PN B10093AE
2-7
Operation
Manual Operation
2
The
RPM = - -
G
field flashes.
3
Enter the required RCF on the keypad and press
ENTER
.
•
The RCF and the corresponding RPM appear in the Speed: field.
•
The RCF (
g
) indicator illuminates on the actual value display, indicating that RCF (
g
) is
being displayed during the run.
•
When
ENTER
and
START
are pressed, the centrifuge runs at the calculated RPM.
•
If you enter an RCF value that is unachievable for the installed rotor, a message will appear
telling you the rotor’s achievable RCF.
Entering Run Time
Run time can be entered in hours and minutes or in
2
t
(accumulated centrifugal effect). Hold mode
is used for continuous operation.
Entering a Timed Run
1
Press
TIME
.
HH:MM:
,
W2T,
and
HOLD
soft keys appear.
2
Use they keypad to enter the required time (up to 99 hours, 59 minutes) and press
ENTER
.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
ROTOR ID
SPEED
RPM/RCF
TIME
HR:MIN
°C
TEMP
Rotor : JA-30.50
Speed : - - RPM = 81 800 x G
Time : 00:00
Temp : 25°C Max Temp : 29°C
Accel : MAX Decel :
RPM
RCF
RCF (x g)
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......