COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
4-61
Operating Instructions
Changing the Dose During an Infusion
Changing the Dose During an Infusion
1.
While the infusion is running:
The
Dose
field is highlighted automatically (Figure 4-87).
2.
Enter the new dose using the numeric keypad. The pump
automatically calculates the new infusion rate (Figure 4-88).
3.
Check the programmed values, then press the
Confirm Primary
soft
key.
4.
Press the
START
key to begin the infusion using the new dose. If the
START
key is not pressed, the pump continues infusing at the
previous dose.
Changing Volume, Weight or Concentration During an Infusion
Use this procedure to change volume to be infused, patient weight, or
concentration of the infusion currently shown on the display.
Figure 4-87 Primary Programming
Screen — Infusion Running
For single channel pumps:
For triple channel pumps:
press the
Primary
soft key.
press the desired
Channel
Select
key.
Note:
Changing
Primary Program
Alert
While the new dose is being entered, a
Changing
Primary Program
alert is active. The alert will be
cleared when the infusion is started with the new rate.
Figure 4-88 Enter New Dose
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...