071960491
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
11-21
Quick Reference Guide
Troubleshooting Alarms
4.
If the tube loading mechanism is open,
close the regulating clamp
on the administration set
and remove the set. Press the
Done
soft
key. The channel is now shown as
Out Of Service.
For single channel pumps: the pump cannot be used to deliver
infusions, but the
Volume
History
key can be used to retrieve history
5.
If the channel fails with the tube loading mechanism in the “closed”
position:
Close the regulating clamp on the administration set.
Press the
Open
key. If the mechanism opens, remove the set.
If the mechanism does not open, use the MTR to remove the set.
6.
Press the
Done
soft key. The Main Display shows that the pump is
out of service.
7.
Do one of the following:
For single channel pumps, cycle power on the pump by powering
off and then powering on again. Do this only once. Proceed to
step 8.
For triple channel pumps, allow any infusions running on the
other pump channels to complete. Remove the pump from service
and have it inspected by Baxter-trained, qualified personnel as
soon as possible. Do not continue with this procedure.
8.
Based on the result, do one of the following:
If the failure code recurs after the pump is turned back on, stop
using the pump.
If the failure code does not recur after the pump is turned back on,
re-load the administration set into the same channel, open the
regulating clamp, and continue using the pump.
9.
Monitor the pump until replacement pump arrives and transfer any
infusions to the replacement pump as soon as it is clinically safe.
Have the failed pump serviced as soon as possible.
Troubleshooting Alarms
An alarm will override an existing alert condition. The alarm tone can be
silenced for two minutes by pressing the
Alarm Silence
key.
Air Detected
Pump module display message:
AIR
Cause:
An air bubble detected in the administration set.
Action:
For single channel pumps: proceed directly to step 2.
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...