4
Primary Infusions
4-40
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
Programming a Primary Dose Mode Infusion (Weight-Based)
1.
From the Main Display, access the Programming screen:
2.
From the Programming screen, press the
Change Mode
soft key.
The Programming Modes menu is displayed (Figure 4-58).
3.
Use the
×Ø
and/or
Page Up, Page Down
soft keys to highlight the
appropriate dose mode selection (Figure 4-59). Some or all of the
following dose modes may be available depending on pump
configuration at the facility:
•
mcg/kg/min
•
mg/kg/hr
•
mEq/kg/hr
•
mg/kg/min
•
mcg/kg/hr
•
units/kg/hr
For single channel pumps:
For triple channel pumps:
press the
Primary
soft key or
the
Rate
or
Volume
keys.
press the desired
Channel
Select
key.
Figure 4-58 Programming Modes
Figure 4-59 Weight-Based Dose
Programming Modes
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...