COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
8-11
Troubleshooting
Troubleshooting Alarms
When the pump detects fluid,
FLUID
3.
Release the
Advance Air soft
key. The pump stops infusing (Figure
4.
Press the
Done
soft key. The
Air Detected
alarm is reset. Visually
inspect the air and follow the facility’s procedures for manually
removing the air.
5.
After the air has been removed, the infusion may be restarted. Press
the
Primary
or
Secondary
soft key to access the appropriate
programming screen and then press the
START
key.
Figure 8-11 Fluid Detected
Figure 8-12 Fluid Detected,
Advance Air Stopped