4
Primary Infusions
4-38
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
7.
Enter the desired Diluent Volume using the numeric keypad.
8.
Use the
Ø
key to highlight the
Dose
field.
9.
Enter the desired dose using the numeric keypad.
If desired, the
Dose
field can be bypassed by using the
Ø
key and
entering the
Rate
value first. The pump then calculates the Dose.
10.
Use the
Ø
key to highlight the
Volume to be infused
field (Figure
The
Volume to be infused
field defaults to the Diluent Volume.
The Volume to be infused may be less than the Diluent Volume, but
cannot be greater than the Diluent Volume.
11.
If appropriate, change the
Volume to be infused
value.
Figure 4-54 Value High Prompt
Note:
Concentration
Too High
The pump calculates the concentration as diluent vol-
ume is entered. In Figure 4-54, the concentration is too
high because the pump is calculating as the numbers
are being entered. This
High
condition is cleared after
the rest of the digits are entered and the pump recalcu-
lates the concentration.
Figure 4-55 Enter Volume to be
Infused
Note:
Optional Labels
An optional label may also be selected. See “Selecting
a Label” on page 4-23 for details.
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...