Quick Reference Guide
Programming a Secondary Volume-Time Infusion
Programming a Secondary Volume-Time Infusion
1.
For single channel pumps: press the
Secondary
soft key.
For triple channel pumps: press the desired
Channel Select
key,
then press the
Secondary
soft key.
2.
Press the
Change Mode
soft key. The Programming Modes Menu is
displayed.
3.
Highlight
Secondary Volume-Time
, then press the
Select
soft
key. The Volume-Time Programming screen is displayed (Figure
11-5).
4.
Enter the
Volume to be infused
using the keypad.
5.
Highlight
Time Duration
using the
×Ø
keys. Use the keypad to
enter the time period for the infusion in hours and minutes. The
pump automatically calculates the flow rate.
6.
Press the
Confirm Secondary
soft key.
7.
Press the
START
key to start the infusion. Confirm that flow is
occurring by observing drops falling into the drip chamber.
Standby Mode
For single channel pumps:
1.
Ensure the pump is stopped.
2.
(Optional) To preprogram the pump for future use, program the
infusion (but do not press the
START
key).
3.
From the Programming screen, press the
Change Mode
soft key.
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...