4
Using the Optional Prime Function
4-18
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
8.
For triple channel pumps only: Arrange the tubing in the tubing
guide according to pump channel.
Using the Optional Prime Function
Overview
The optional Prime function can be used to assist clinicians in preparing
a primary administration set for infusion.
If Prime is not displayed on the Programming Modes menu, this feature
has not been enabled.
Priming the Administration Set
Note:
The administration set’s drip chamber should be at least one third
full prior to using the prime function to ensure that fluid will enter
the administration set.
Note:
Prime cannot be selected if an
Air
alarm is active.
Note:
Air detection is disabled when priming is active.
! WARNING !
If flow is observed when tubing is loaded but the pump is not running, close the
regulating clamp immediately. Ensure that all steps have been properly
performed. If flow is still observed, remove the pump from service and contact
Baxter-trained, qualified personnel.
! WARNING !
Do not connect the administration set to the patient when priming.
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...