071960491
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
11-9
Quick Reference Guide
Programming a COLLEAGUE GUARDIAN Infusion (Rate-Volume)
1.
For single channel pumps: press the
Primary
soft key.
For triple channel pumps: press the desired
Channel Select
key.
2.
Press the
Colleague Guardian
soft key from the Programming
screen. The labels for which COLLEAGUE GUARDIAN limits
have been defined are displayed in a pop-up window (Figure 11-6).
3.
Use the
×Ø
keys to highlight the desired label, then press the
Select
soft key.
The programming mode changes to the mode configured for the
selected label, and the Rate field is filled with the configured value
4.
Use the
Ø
key to highlight the The
Volume To Be Infused
field.
5.
Enter the volume using the numeric keypad.
6.
Press the
Confirm Primary
soft key. If the values entered result in a
dose that is outside the COLLEAGUE GUARDIAN rate limits, a
Limits Warning pop-up is displayed (Figure 11-7).
If this occurs, do one of the following:
Press
Cancel Rate
(
Ø
key) to return to the programming screen,
then enter a rate that is within the rate limits.
Figure 11-6 Colleague Guardian
Label List
Figure 11-7 Limits Warning Pop-up
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...