Display Reference Guide
2-14
COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL
071960491
As the battery charge level continues to decrease, the battery charge icon continues to
change. When the battery time remaining is 20 to 60%, the icon contains two yellow
bars as shown at left.
When the battery time remaining is less than 20%, the battery charge icon contains
one red bar as shown at left.
When the battery time remaining is only 5 minutes, the battery charge icon appears as
an empty outline as shown at left. The icon will remain an empty outline until the
battery is depleted and pump shuts itself off.
If the pump’s batteries need replacement when the pump is first powered on, a
Damaged Battery! Service Now
alarm
occurs and this icon flashes on the
Power On screen. The pump cannot be used. Send the pump to service.
If the pump’s batteries need replacement during pump operation, a
Damaged Battery! Service Now
alert
occurs and this icon flashes instead of
the battery charge icon. Ensure the pump is plugged into AC power. Do not use the
pump for transport. Have the pump serviced as soon as possible so the batteries can be
replaced.
When the pump is plugged in, the battery charge icon alternates with the charging icon
shown at left.
When the pump is unplugged and operating on battery power, the battery charge icon
alternates with the Plug In icon shown at left, and the approximate battery operating
time is displayed below the icon. The pump should be plugged in whenever possible
to maintain battery charge.
A drop icon is displayed for each channel that is running, and beside the FLUID and
AIR icons when the pump is used to purge air from the tubing following an Air In
Line alarm.
The flow check icon indicates approximate level of downstream occlusion. The
greater the resistance to fluid flow, the greater the number of solid arrows displayed.
This icon indicates the screen is displaying secondary infusion information for a
channel.
This icon is displayed on the Programming screen when a pump channel is stopped.
This icon is displayed on the Main Display screen when a pump channel is stopped.
The appropriate channel letter is displayed inside of the stop symbol (A, B, or C).
When the keypad is locked, the Lock icon is displayed between the second and third
soft keys. The following keys remain available when the keypad is locked so that
infusion status information can be viewed:
Main Display, Back Light, Volume
History,
Primary
,
Secondary
and
Options
.
Table 2-6
Main Display Icons — continued
Icon
Description
Содержание Colleague 3 CXE
Страница 24: ...1 Indications for Use 1 12 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 50: ...2 Infusion Modes 2 26 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 54: ...3 Check out 3 4 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 152: ...7 Storage 7 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 194: ...9 Electromagnetic Compatibility Statement 9 20 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 220: ...11 Troubleshooting Alerts 11 24 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 228: ...Index 8 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP OPERATOR S MANUAL 071960491 ...
Страница 229: ...071960491 ...