en_BA_PosCon_OXE7.docx
7/75
Baumer Electric AG
04.01.2021 15:33/adr V1.3 ANW_81125407
Frauenfeld, Switzerland
2.3
Measuring mode
Edge: Edge position (basic setting)
To perform edge measurement, select FUNCTION
"EDGE" in the menu. Within EDGE the edge to be
measured is defined in MEASUREMENT TYPE.
EDGE L RISE
= First rising edge from the left
EDGE L FALL
= First falling edge from the left
EDGE R RISE
= First rising edge from the right
EDGE R FALL
= First falling edge from the right
EDGE L RISE EDGE L FALL
EDGE R RISE EDGE R FALL
EDGE HEIGHT
Minimum height to be detected as an edge.
OBJECT
Selection of light or dark objects to optimize the
measurement results.
PRECISION
For more reliable measurement results, the output
values can be filtered with High and Very High.
FLEX MOUNT
If the sensor is mounted at an angle, FLEX
MOUNT must be activated and the reference
surface must be taught in.
1. Activate FLEX MOUNT
2. Align and confirm sensor or reference
surface
3. When all conditions are met (see table on
right), confirm by pressing Set for 2 seconds
4. Enter thickness of the auxiliary plate (if
present)
Distance between sensor and
reference surface
Installation angle too large
Reference surface too uneven
Reference surface too small
(<50 mm)
3a
Min. edge height
Object
Reference
surface
Содержание PosCon OXE7
Страница 1: ...Operating instructions PosCon OXE7 Edge measurement in a new dimension...
Страница 27: ...en_BA_PosCon_OXE7 docx 27 75 Baumer Electric AG 04 01 2021 15 33 adr V1 3 ANW_81125407 Frauenfeld Switzerland...
Страница 52: ...6 2 Dimensions Optical axis...
Страница 75: ...en_BA_PosCon_OXE7 docx 75 75 Baumer Electric AG 04 01 2021 15 33 adr V1 3 ANW_81125407 Frauenfeld Switzerland...