
en_BA_PosCon_OXE7.docx
37/75
Baumer Electric AG
04.01.2021 15:33/adr V1.3 ANW_81125407
Frauenfeld, Switzerland
5.8.4
THICKNESS REF
In this menu item, the reference surface is finally defined with due regard to the thickness of the auxiliary
plate.
Important: The sensor automatically suppresses everything below the defined reference surface. This
reference surface displacement will also be used to suppress unwanted backgrounds.
The surface taught in under TEACH REF is always the basis for this. This can be corrected downward with a
positive value.
TEACH REF; taught-in surface
10 mm (example)
THICKNESS REF = 10 mm (example); this is the redefined reference surface, everything
below it is suppressed
NOTE
If an auxiliary plate is not used, the item THICKNESS REF must be saved with 0
mm by pressing SET for 2 seconds.
Auxiliary plate
NOTE
When FLEX MOUNT is activated, the measuring field and the digital switching
output are reset to the standard setting (FLEX MOUNT = Maximum measuring
field, DIGITAL OUT = 0 mm).
Содержание PosCon OXE7
Страница 1: ...Operating instructions PosCon OXE7 Edge measurement in a new dimension...
Страница 27: ...en_BA_PosCon_OXE7 docx 27 75 Baumer Electric AG 04 01 2021 15 33 adr V1 3 ANW_81125407 Frauenfeld Switzerland...
Страница 52: ...6 2 Dimensions Optical axis...
Страница 75: ...en_BA_PosCon_OXE7 docx 75 75 Baumer Electric AG 04 01 2021 15 33 adr V1 3 ANW_81125407 Frauenfeld Switzerland...