Problem
Possible Causes
Solutions
EASYpure RO
No electrical power to
Ensure EASYpure RO
completely inactive.
EASYpure RO.
power cord is installed
(Controls not operating.)
and connected to live power.
Blown fuses.
Check/Replace fuses.
Refer to “Fuse Replacement”
for procedure.
No water flowing (low
Feedwater supply valve
Open feedwater supply valve.
pressure indicator light
illuminated).
Prefilter plugged.
Replace prefilter.
Solenoid valve not open.
Replace solenoid valve.
Feed pressure below
Increase pressure.
specifications.
Illumination of service
Improperly installed
Ensure membrane is
membrane light.
membrane.
installed properly.
Decrease in % rejection.
Fouled membrane.
Replace membrane.
Decrease in system
Decrease in water
If decline in flowrate is
productivity.
unacceptable, install
feedwater temperature valve
to elevate water temperature.
Membrane fouled.
Replace membrane
Low incoming water
Ensure prefilter is not
pressure.
blocked—increase incoming
water pressure.
Low membrane
Increase operating pressure.
operating pressure.
32
Troubleshooting