Electrical Connections
1.
Plug power cord into the unit's power
entry module.
2.
Connect your EASYpure RO to a prop-
erly grounded outlet of correct voltage
and current handling capacity.
Membrane Flush
1.
If not already done, disconnect the prod-
uct water tubing from any EASYpure
unit connected to your EASYpure RO.
Direct the product water tubing to drain.
If you are feeding an accessory 30 Liter
Reservoir, disconnect the product water
tubing from the accessory 30 Liter
Reservoir. Direct the product water tub-
ing to drain.
2.
Disconnect the pump interlock cord from
the other EASYpure unit.
3.
Open the customer-supplied feed water
valve.
4.
Turn the unit on by depressing the
ON/OFF switch on the power entry
module in the rear of the EASYpure RO.
(See Figure 8)
5.
Allow the unit to operate for
approximately one hour, sending all
water produced to drain.
16
I
NSTALLATION
Warning
Use a properly grounded electrical out-
let of correct voltage and current han-
dling capacity.
Figure 8: Power Cord Adjustment
Power Entry
Module