4.
Disconnect the quick disconnect
insert(s) and product water tubing
attached to the EASYpure RO.
5.
Remove the pretreatment cartridge
and membrane cartridges.
6.
Install the membrane bypass tubing
on the top membrane connectors.
7.
Install the bottom bypass into the two
membrane inlets.
a.
Wet the o-rings on the connec-
tors of the bypass tubing.
b.
Insert the connectors into the
two membrane inlets. Ensure
that the metal clamps line up
with the slots located on the
floor of the cabinet.
c.
Rotate the connectors to lock
the metal clamps into position.
8.
Install the sanitization cartridge in the
pretreatment cartridge position.
9.
Power up the EASYpure RO and
allow the water to move from the inlet
line, through the sanitization cartridge,
and through the rest of the system.
The unit may turn off as a result of the
low pressure condition once or twice
during this step. If this occurs, power
down and power back up again.
10. Allow the system to operate normally
until the reservoir is full and the unit
goes into standby.
23
M
AINTENANCE