Reject Water Connection
1.
Locate the reject water tubing. This is
the piece of 3/8" O.D. tubing that has
an adapter on one end and no con-
nector on the other end. Remove the
nut, grab ring, backup ring and o-ring
from the reject water fitting located at
the lower right of the unit. (See Figure
4 for part identification and Figure 3
for fitting location.)
2.
Thread the nut, grab ring and backup
ring on the reject adapter. Do not use
the o-ring at this time.
3.
Push the tubing through the hex nut
until it bottoms out in the adapter.
4.
Remove the adapter nut and tubing.
Place the o-ring over the tubing. Be
careful not to push the backup ring or
grab ring further back on the tubing
when installing the o-ring. Connect
the tubing to the reject connection on
the EASYpure RO.
5.
Route the other end of the reject
water tubing with the adapter installed
to an atmospherically vented drain.
Ensure there are no kinks in the tub-
ing and that it proceeds in a
downward plane.
6.
If possible, use the tubing adapter to
permanently install the reject water
tubing in the drain.
11
I
NSTALLATION
Power Entry
Module/
ON-OFF
Switch
Product Water Outlets
Pump
Interloc
k
Tank
Overflow
Reject Water
Connection
Feedwater
Inlet
Figure 3: Connections
Figure 4: Typical Polypropylene Tubing Adapter Installation