6.
Turn unit power off. Reconnect the
product water tubing to the other
EASYpure unit. Reconnect the product
water tubing to the accessory 30 Liter
Reservoir. Reconnect the pump inter-
lock cord to the other EASYpure unit.
Set "Replace Prefilter" Indicator
The extruded carbon prefilter both filters partic-
ulates and absorbs free chlorine out of your
feedwater supply. After a certain point, the
extruded carbon will become saturated with
chlorine and will cease to effectively protect
your membrane from the damaging effects of
chlorine. The "Replace Prefilter" Indicator alerts
you when the prefilter requires changing.
To set the "Replace Prefilter" Indicator for your
feedwater supply:
1.
Determine the free-chlorine level of your
feedwater (in ppm) using a commercial-
ly available test kit.
2.
Disconnect the unit from the power sup-
ply.
3.
Pull the rear side of the door latch up
and toward the front of the unit until the
front of the door latch unhooks.
4.
Lift the front of the door latch and open
door.
5.
Locate the switch panel on the upper
left sidewall of the membrane compart-
ment. (See Figure 9 for location.)
6. The four rocker switches on the switch
panel are numbered from left to right as
17
I
NSTALLATION
Figure 9 Switch Panel Location
Membrane
Bypass
Tubing
Switch
Panel
Prefilter Reset
Membrane
Reset