General Cleaning Instructions
Wipe exterior surfaces with lightly dampened
cloth containing mild soap solution.
Normal Unit Operation
Below is a brief description of the normal auto-
matic operation of the EASYpure RO. If the unit
does not operate as described below, refer to
the Troubleshooting Guide.
When you activate the power switch, the
following events will occur:
1.
All LEDs light momentarily to indicate
that they work.
2.
After approximately 5 seconds, the
automatic solenoid valves are ener-
gized. The inlet valve opens, supplying
the EASYpure RO with water. The flush
valve also opens, diverting virtually all
the water to drain.
3.
After approximately 20 seconds, the
pump energizes. The majority of the
water produced goes to drain.
4.
10 seconds after the pump energizes,
the pressure switch begins monitoring
the incoming water pressure.
5.
After 3 minutes, the flush solenoid valve
closes and the EASYpure RO begins
making product water and sending it to
the storage reservoir.
6.
The unit begins to monitor membrane
performance and the storage reservoir
floats begin to monitor tank levels 5 sec-
26
Service and Repair
Warning
Disconnect from the power supply
prior to maintenance and servicing.
Refer servicing to qualified personnel.
Ensure all piping connections are tight
to avoid leakage of chemicals.
Ensure adequate ventilation when
using chemicals for
cleaning.
Follow carefully the manufacturer’s
safety instructions on labels of chemi-
cal containers and Material Safety
Data Sheet (M.S.D.S.)