Tank Overflow Connection
1.
Refer to Figure 3 and attach a customer
supplied hose barb connector and 5⁄8"
O.D. tubing to the tubing located at the
lower right of the unit as you look at the
rear of the unit.
2.
Route tubing to an atmospherically vent-
ed drain. Ensure there are no kinks in
the tubing and that it proceeds in a
downward plane.
Connections to Equipment
Supplied By the EASYpure RO.
Your EASYpure RO also includes two lengths of
tubing with quick disconnect inserts attached.
One is 1/4" O.D. tubing and one is 3/8" O.D.
tubing. Use the 1/4" tubing to feed directly from
the EASYpure RO to another EASYpure unit.
Use the 3/8" O.D. tubing to feed the accessory
30 Liter Reservoir if one is being used.
1.
Install the quick disconnect insert
attached to the tubing into one of the
product water quick disconnect
couplings on the EASYpure RO.
2.
Install the other end of the tubing into
the unit being fed by the EASYpure RO
according to the instructions in that
unit's manual.
3.
Locate the pump interlock cord supplied
with the EASYpure RO. Plug one end of
the pump interlock cord into the pump
interlock on the EASYpure RO. (See
Figure 3.) Remove the jumper plug from
the pump interlock on the other
12
I
NSTALLATION
Note
Do not connect the product water tub-
ing to a connected EASYpure unit until
you have completed the membrane
flush procedure for the EASYpure RO.
Note
The product water quick disconnect
inserts can be installed in either prod-
uct water quick disconnect coupling.
The quick disconnect ends of the
inserts are the same size.
Note
If you are not now feeding another
EASYpure unit, save the pump inter-
lock cord for possible future use.