Fuse Replacement
1.
Disconnect the EASYpure RO from its
power source. Remove the outer cover.
2.
Locate the fuse holder, part of the
power entry module (see Figure 13 for
location), and remove.
3.
Remove the burned out fuse(s) and
replace with one(s) of equal size and
rating.
Membrane Replacement
A variety of factors will determine the service life
of your EASYpure RO membranes. With proper
maintenance, the membranes should last
approximately one year for normal usage.
When either the "Service Membrane" or the
"Replace Membrane" Indicator lights, use the
following procedure to replace the membranes:
1.
Disconnect the EASYpure RO from its
power source.
2.
Remove the membranes and discard.
3.
Install and flush new membranes using
the procedure for membrane installation
and flush listed under "Membrane
Installation."
4.
After you replace the membranes, press
the "Replace Membrane" Indicator reset
button, located directly below the
"Replace Prefilter" Indicator reset but-
ton, which is in turn directly below the
switch panel on the upper left sidewall
of the membrane compartment. (See
Figure 9.)
28
S
ERVICE AND
R
EPAIR
Warning
Replace fuses with those of the same
type and rating.
Figure 13: Fuse Holder Location
Note
The RESET button is a two
position button; up or down. Press it
once to change its position. (If it was
up before you changed the mem-
branes, press it down after you change
the membranes and vice versa.
Fuse Holder
Location