Your EASYpure RO was designed to operate au-
tomatically. The unit will automatically turn off
when the tank is full and will restart when the
tank empties slightly. The unit will restart auto-
matically for 5 minutes every 3 hours 55 minutes
of inactivity to flush the membrane during periods
of non-use. This flush water will automatically run
to the drain. Please ensure that the unit remains
on — if you need to de-energize the system for
longer then 48 hours contact Barnstead
International customer service. If water is not
drawn from the tank for 72 hours, empty tank
and refill with new RO water before using.
Indicators
(See Figure 11.)
Pressure Gauge
On initial start-up, the pressure gauge will briefly
indicate the feedwater pressure. Once the pump
starts, the pressure gauge will indicate the pres-
sure after the pump, which is the pressure on the
reverse osmosis membranes. During normal
operation, this gauge should read 55-65 psi.
LED's
There are five indicator lights on the EASYpure
RO front panel. Their significance is as follows:
Indicator Lights:
These lights alert you of the
operating status of the unit. If your attention is re-
quired to correct a problem or maintain the sys-
tem, it will be indicated on the display (periodical-
ly check the panel to see if attention is required).
A listing of the various indicator lights and their
meanings follow:
1.
Reservoir full
: Your EASYpure RO is
equipped with a float switch that will de-
energize the unit when the reservoir is
19
Operation
Warning
Do not use in the presence of flam-
mable or combustible materials; fire
or explosion may result. This device
contains components which may
ignite such materials.