the end cap slide down on each side of
the keyway wall. (See Figure 7.)
4.
Lower the prefilter and insert the lower
end cap into the lower center socket
until it is firmly seated.
Membrane Installation
1.
Remove the membrane bypass tubing
from the EASYpure RO (see Figure 5).
The membrane bypass tubing may be
stored inside the EASYpure RO case for
easy access when needed for the
system sanitization procedure.
2.
Remove an RO membrane cartridge
(Part Number D7425) from its plastic
bag.
3.
Remove the nut, stainless steel grab
ring, backup ring and o-ring from the
connectors on the membrane.
4.
Connect the product and reject tubes in
the right-hand membrane position on
the EASYpure RO to the product and
reject outlet connectors on the RO
membrane cartridge. Ensure that the o-
ring, backup ring, stainless steel grab
ring and nut are in place on the product
and reject tubing on the EASYpure RO
(see Figure 4). Ensure that the product
tube is connected to the product
connector and the reject tube is con-
nected to the reject connector. Hand-
tighten the connections.
14
I
NSTALLATION
Figure 5: Membrane Bypass Tubing Installed
Note
The membrane by-pass tubing will be
needed for sanitizing the EASYpure
RO. DO NOT DISCARD IT.
Membrane By-Pass
Tubing