WirelessHART tri-axial
sensor (70M403)
Detects velocity and acceleration in three axis (X, Y, and Z) and
measures surface temperature.
ISA100 tri-axial sensor
(70M303)
Detects velocity and acceleration in three axis (X, Y, and Z) and
measures surface temperature.
ISA100 single axis
sensor (70M301)
Detects velocity and acceleration in one-axis (vertical, or Z) only and
measures surface temperature.
ISA100 repeater
(70M300)
Enables you to extend the range between sensors and network access
points.
For ISA100 networks, both the tri-axial and uniaxial sensors can also act as routers, although
using the router mode depletes the unit’s battery more quickly.
If a Ranger Pro device can’t communicate with an backbone router, you can use the Ranger Pro
ISA100 repeater or a Ranger Pro ISA100 device enabled as a router to extend your network. We
recommend connecting no more than three Ranger Pro sensors to a Ranger Pro router,
although you may be able to connect up to eight.
8/56
Ranger Pro Wireless Condition Monitoring Device
User Guide
125M6113 Rev. D