Azbil Corporation
Repair
Model SLX 110/120 - Smart displacement type level transmitter 4-7
Message
Meaning
Action
CORRECT LRV ?
Is the input value appropriate
for the value to be set?
Check the input value. If
wrong, correct it.
ENTRY > SENS RNG
When the set value exceeds
the allowable setting range
of this instrument
Press [CLR], check the value
and set it again.
CLR
( No )
INVALID REQUEST
This request cannot be
executed.
Check the SFC operation
procedure.