![Azbil SLX 110 Скачать руководство пользователя страница 134](http://html1.mh-extra.com/html/azbil/slx-110/slx-110_user-manual_3042728134.webp)
Maintenance
Azbil Corporation
3-14
Model SLX 110/120 - Smart displacement type level transmitter
Liquid gauge expression
Fb
: Mass of float (FLOAT LOAD is given on identification plate or is the measured value of
float mass)
D
: Float diameter (mm)
H
: Float length (mm)
: Specific gravity (g/cm
3
)
a
: Constant,
a
= 5.76 x 10-7
Interface gauge expression
Fb
: Mass of float (FLOAT LOAD described on identification plate or measured value of float
mass)
r
1
: Specific gravity of upper layer liquid (g/cm3)
r
2
: Specific gravity of lower layer liquid (g/cm3),
r
2
>
r
1
Other parameters are same to (a) liquid gauge.
3.
Turn the power to the instrument ON and connect the SFC.
(See “2-3 : Items to be checked before operation” on page 2-5.)
4.
Make sure the value of specific gravity conforms with the specific gravity of the
fluid to be measured and carry out the setting.
(See “Displaying and changing set data for measurement” on page 2-37.)
5.
Calibrate the LRV.
Step
Button
Procedure
SFC screen
1
Apply 0% calibrated load to the torque
arm.
SLX TAG - 0001
READY...
2
LRV E
0%
Press [LRV]. The current LRV will
appear.
Example: 0.12 mm
LRV TAG - 0001
0.1200 mm
Содержание SLX 110
Страница 1: ...Smart Displacement type Level Transmitter Model SLX 110 120 User s Manual CM2 SLX100 2001...
Страница 4: ......
Страница 8: ...iv Model SLX 110 120 Smart displacement type level transmitter...
Страница 40: ...Installation Azbil Corporation 1 26 Model SLX 110 120 Smart displacement type level transmitter...
Страница 120: ...Operation Azbil Corporation 2 80 Model SLX 110 120 Smart displacement type level transmitter...
Страница 150: ...Maintenance Azbil Corporation 3 30 Model SLX 110 120 Smart displacement type level transmitter...
Страница 162: ...Azbil Corporation 5 2 Model SLX 110 120 Smart displacement type level transmitter...
Страница 163: ......
Страница 166: ......
Страница 168: ......