Azbil Corporation
Maintenance
Model SLX 110/120 - Smart displacement type level transmitter 3-9
(*) Read from the SFC printout result.
~Note
Read
D
PV after making sure the output of this instrument has been
stabilized.
3. If
satisfies expression (1), this means that the torque tube temperature
compensation functions effectively, and therefore it is recommended to select zero
adjustment (pinpoint adjustment) that adjusts to the level in
j
.
4. If
does not satisfy expression (1), it is recommended to select the change to
the torque tube temperature coefficient in
k
. In this case, calculate the torque tube
temperature coefficient to be changed from expression (2) and set the coefficient
using the SFC.
............
(2)
: Torque tube temperature coefficient to be changed
: Current torque tube temperature coefficient (*)
: Material coefficient (*)
: Input angle (*)
: Torque tube temperature (°C) (*)
: Angle sensor temperature (°C) (*)
(*) Read from the SFC printout result.
5.
After confirming that the output of the instrument has been stabilized, read
D
PV
and if
satisfies expression (1), conduct a zero adjustment that adjusts the level
in
j
(pinpoint adjustment).If
D
PV does not satisfy expression (1), change the
torque tube temperature coefficient until the output of the instrument matches the
reading of the glass gauge.
At this time, estimate the torque tube temperature coefficient to be changed from
the correlation between past torque tube temperature coefficients and
D
PV.
The following is the method of setting a torque tube temperature coefficient using the
SFC.
Step
Button
Procedure
SFC screen
1
B
CONF
Press [CONF].
SLX CONFIG
FUNCTION CONFIG ?
Содержание SLX 110
Страница 1: ...Smart Displacement type Level Transmitter Model SLX 110 120 User s Manual CM2 SLX100 2001...
Страница 4: ......
Страница 8: ...iv Model SLX 110 120 Smart displacement type level transmitter...
Страница 40: ...Installation Azbil Corporation 1 26 Model SLX 110 120 Smart displacement type level transmitter...
Страница 120: ...Operation Azbil Corporation 2 80 Model SLX 110 120 Smart displacement type level transmitter...
Страница 150: ...Maintenance Azbil Corporation 3 30 Model SLX 110 120 Smart displacement type level transmitter...
Страница 162: ...Azbil Corporation 5 2 Model SLX 110 120 Smart displacement type level transmitter...
Страница 163: ......
Страница 166: ......
Страница 168: ......