6-34
No.
Error message
Checkpoint and action
17
RAM FAULT
A critical failure. Stop the electromagnetic flow
meter.
The RAM of the converter is abnormal. Turn OFF
the power and then turn it ON again and check the
operation.
If the same message still appears after taking the
action above, contact an Azbil Corp. representative.
18
ROM FAULT
A critical failure. Stop the electromagnetic flow
meter.
The ROM of the converter is abnormal. Turn OFF
the power and then turn it ON again and check the
operation.
If the same message still appears after taking the
action above, contact an Azbil Corp. representative.
19
SFC FAULT
An SFC error.
Replace the SFC.
20
SPAN OVER ERROR
As a result of setting the span, the maximum
measurable flow rate has exceeded 12 m/s.
Check the span, diameter or type of the detector.
21
>RANGE
The calculation result of the SFC has exceeded the
display range.
Restart the SFC.
The SFC battery is running low.
22
:
Charge the SFC.
A minor fault.
23
#
Press the
U
STAT
F/S DIR
key and check the self-diagnostics
result of the SFC.