6-69
Step
Key
Procedure
SFC screen
5
MENU I
ITEM
DE CONF.
Display shown to the right appears.
Press the
MENU I
ITEM
DE CONF.
key and select the
value to be calibrated.
The value to gain calibrate for two wired
magnetic flowmeter are the following
three: 0 m/s, 2.5 m/s, and 10 m/s. Value
2.5 m/s for two wired magnetic
flowmeter will be shown 1.2 m/s in SFC.
When calibrating to 2.5 m/s select 1.2
m/s.
SFCM00006125D
SFCM00006126D
6
ENTER
( Yes )
NON-VOL
For example, when calibrating 2.5 m/s,
displays screen as shown to the right.
For smart calibrator model MGZ14,
enter 2.5 m/s.
Then, press the
ENTER
( Yes )
NON-VOL
key. Gain
calibration starts. When gain calibration
completes, display will return back to the
screen to select the gain calibration
value.
SFCM00006126D
SFCM00006126D
SFCM00006128D
7
Follow the same procedure for
calibrating 0 m/s and 10 m/s.
8
MENU I
ITEM
DE CONF.
When gain calibration is completed,
press the
ENTER
( Yes )
NON-VOL
key and the display as
shown on the right appears.
SFCM00006127D
9
ENTER
( Yes )
NON-VOL
Press the
ENTER
( Yes )
NON-VOL
key.
Exits from gain calibration function.
SFCM00006128D
SFCM00006129D
10
CLR
( No )
Press the
CLR
( No )
key and the screen
returns to the screen as in step 1.
SFCM00006009D
Содержание MagneW Two-wire PLUS+ MTG15A
Страница 8: ...vi...
Страница 22: ...1 14...
Страница 56: ...2 34...
Страница 60: ...3 4 Fig 3 2...
Страница 66: ...3 10...
Страница 70: ...4 4...
Страница 144: ...5 74...
Страница 226: ...6 82...
Страница 268: ...7 42...
Страница 274: ...8 6...
Страница 275: ......
Страница 278: ......
Страница 280: ......