5-68
Note) If the main display shows -2.0%, the zero point value may exceed -2.0%. Execute Auto
zero before manual zero.
Step
Procedure
Screen
5
Adjust zero point by pushing the
or
key so that
the main display shows 0.0%.
By pushing the
key once, READY changes to UP
and the zero point value increases 0.05%.
By pushing the
key once, READY changes to
DOWN and the zero point value decreases 0.05%.
It takes about 20 seconds to change the zero point
value. During manual zeroing, the value in the main
display flashes. If you push the
or
key again, wait
until the value in the main display does not flash.
Note) Do not keep on pushing the
or
key.
Manual zeroing does not work.
0.0 %
> MANUAL ZERO1
READY
0.0 %
> MANUAL ZERO1
UP
0.0 %
> MANUAL ZERO1
DOWN
6
Move the cursor to the mode indicator by
pushing
key.
0.0 %
> MANUAL ZERO1
READY
7
Push the
key and display the MANUAL ZERO2
screen.
0.5 %
> MANUAL ZERO2
WORKING
8
Execute MANUAL ZERO2, as well as MANUAL
ZERO1. (Refer to the Step 2 to 6.)
0.0 %
> MANUAL ZERO2
WORKING
9
Push the
key and display the MANUAL ZERO3
screen.
0.5 %
> MANUAL ZERO3
WORKING
10
Execute MANUAL ZERO3, as well as MANUAL
ZERO1. (Refer to the Step 2 to 6.)
0.0 %
> MANUAL ZERO3
WORKING
11
Push the MODE key and return to the MEASURING
MODE.
0.5 %
MEASURING MODE
Содержание MagneW Two-wire PLUS+ MTG15A
Страница 8: ...vi...
Страница 22: ...1 14...
Страница 56: ...2 34...
Страница 60: ...3 4 Fig 3 2...
Страница 66: ...3 10...
Страница 70: ...4 4...
Страница 144: ...5 74...
Страница 226: ...6 82...
Страница 268: ...7 42...
Страница 274: ...8 6...
Страница 275: ......
Страница 278: ......
Страница 280: ......