Non-critical error
Non-critical failures will not seriously affect electromagnetic flowmeter operation.
When an error occurs during operation and is regarded as a non-critical problem by the
converter self-diagnostics, the output will not burn-out and the electromagnetic flowmeter
will continue to output the measured value.
If a wrong setting is found, an error message is displayed for a second, and then the screen set
wrongly is displayed.
Error code of set errors
Error
code
Error content
Action
LCD display
Err-12
Upper/lower limit alarm
set error
HI < LO is set.
Set HI>LO.
Err - 12
SETTING ERROR
HI<LO
Err-21
Span is set to 12 m/s or
more.
Check the settings of
flow rate range and
detector information
(bore diameter and
detector type).
Err - 21
SPAN ERROR
OVER
12 m/s
Err-22
Pulse frequency is too
large or too small.
The flow rate range unit
system is different from
the pulse unit system.
Example: SPAN m
3
/h
pulse scale t/h
1. Check pulse scale.
2. Check the setting of
pulse frequency.
3. Adopt a unified unit
system.
Err - 22
PULSE WEIGHT
SETTING ERROR
Err-23
The pulse width is too
large.
When pulse frequency is
output, the duty is 70%
or more.
Check the following
settings:
1. Pulse width
2. pulse scale
3. Span
Err - 23
PULSE WIDTH
OVER DUTY 70%
Содержание MagneW Two-wire PLUS+ MTG15A
Страница 8: ...vi...
Страница 22: ...1 14...
Страница 56: ...2 34...
Страница 60: ...3 4 Fig 3 2...
Страница 66: ...3 10...
Страница 70: ...4 4...
Страница 144: ...5 74...
Страница 226: ...6 82...
Страница 268: ...7 42...
Страница 274: ...8 6...
Страница 275: ......
Страница 278: ......
Страница 280: ......