4. Click Next to move to the Test Condition screen.
5. Enter the name of the Test Condition. This field contains the name of the condition and
specifications of the test, such as test temperature, stress level, furnace ramp rate and soak time,
type of specimen, and type of test.
6. Use the pull-down menus to select the appropriate Specimen Type and Test Type. Review the
entire list of options to ensure the correct selection is made.
7. Use the Tab key to move through the remaining fields.
8. Units can be changed using the User Units section (see Section 6.1).
9. Note that Ramp Rate is determined at time of installation and is not normally changed unless
testing at an unusually low temperature, when a slower ramp may be necessary to prevent
overshoot as temperature reaches setpoint.