3.5 Hardware Installation
All hardware is built at the ATS factory. If hardware installation is required on site, please contact the
Service Department. ONLY PERFORM THOSE OPERATIONS THAT APPLY TO THE UNIT RECEIVED.
NOTE: ELECTRICAL POWER MUST BE CONNECTED PRIOR TO ARRIVAL OF ATS SERVICE.
3.5.1 Lever Arm Assembly Installation
1. Position the knife-edge support block assembly on top of the test frame so that the serial
number on the block is facing the weights.
2. Install the locating dowels.
3. Install bolts and tighten.
CAUTION: Match the serial numbers on the test frames with the support blocks and lever
arms when assembling multiple testers.
4. Remove or pivot the lever arm safety guards to the side.
5. Check that the knife-edge and V-block mating surfaces are free of dirt or rust.
6. Install the lever arm to the tester with the V-blocks mated to the knife-edges.
NOTE: Take care not to damage knife-edges or proximity switch.
7. Ensure that the lever arm V-blocks are centered on the pedestal block knife-edges front to back.
The lever arm should pivot freely.
8. Install the safety guards and tighten bolts.
Содержание 2140-M Series
Страница 1: ...WinCCS Lever Arm Tester Series 2140 M 2320 M 2330 M 2330 CC 2410 M 2410 CC...
Страница 2: ...Revised March 2013...
Страница 14: ...14...
Страница 95: ...95 Once these steps are completed the thermocouples will be properly attached to the specimen...
Страница 109: ...109 Appendix C Default Values This appendix includes all the default values typically seen on a WinCCS tester...
Страница 110: ...110...
Страница 111: ...111...
Страница 112: ...112...
Страница 113: ...113...
Страница 115: ...154 East Brook Lane Butler PA 16002 USA 1 724 283 1212 www atspa com...