103
A.2 Status Display Colors
Individual frame lines will change color frequently as the frame state and alarm conditions change.
Red:
Major alarm or action required, alarm condition will be displayed.
Green:
Normal run state, test in progress.
Blue:
Indicates normal activity under Temperature Columns and signals a Minor alarm state, alarm
condition will be displayed.
Black:
Column header, idle frame state.
Orange:
Time only, normally indicates time remaining until change of frame state.
A.3 System Status Error Codes
Error codes explain issues with the system. Some of the errors are fatal and will stop the test, while
others supply information and indicate a potential problem but do not indicate that the test is faulty.
ACLow:
The frame controller’s AC power line has dropped below 95VAC, indicating a severe brownout
condition.
ACV:
The frame has no AC power to the frame and/or furnace.
BAT:
The frame controller’s battery for the NVRAM needs replacing.
Break:
The frame’s break detect switch is active.
Cal/Ver:
The frame is currently having the extensometer calibrated or verified.
Com:
The host PC is unable to communicate with the frame controller.
ECnt:
The number of events in the event log is greater than fifty.
FZone:
The furnace zone control algorithm has reached its limits and is unable to control the differential
temperature between zones. This is usually caused by one or all of the following: open furnace zone,
thermocouples in wrong position on the specimen, open zone control relay, or specimen too high or low
in the furnace. See A.3.1 for an extended explanation of the FZone alarm.
LCtrl:
The automatic load control system has faulted. This is usually due to elevator-style machines
running out of travel, and the lever arm must be re-zeroed.
Load:
The test is ready to load weights.
MLO:
Machine Lockout
Open:
A control thermocouple is open.
SErr:
A startup error has occurred. This is caused by a reversed thermocouple or an open thermocouple
specified as active in a test. Once the problem is corrected the test will start automatically.
Содержание 2140-M Series
Страница 1: ...WinCCS Lever Arm Tester Series 2140 M 2320 M 2330 M 2330 CC 2410 M 2410 CC...
Страница 2: ...Revised March 2013...
Страница 14: ...14...
Страница 95: ...95 Once these steps are completed the thermocouples will be properly attached to the specimen...
Страница 109: ...109 Appendix C Default Values This appendix includes all the default values typically seen on a WinCCS tester...
Страница 110: ...110...
Страница 111: ...111...
Страница 112: ...112...
Страница 113: ...113...
Страница 115: ...154 East Brook Lane Butler PA 16002 USA 1 724 283 1212 www atspa com...