B2: States that occur while Running a Test
All Specimen Thermocouples are Open:
All control thermocouples are open and the test will be shut
down.
Automatic Loading Failed:
The automatic creep loading system failed; this statement is usually
accompanied with a reason for load failure.
Automatic Loading Timed Out:
The automatic creep step loading routine failed due to insufficient load
change during the loading procedure.
Cannot Start, Open TCs:
One or more thermocouples are open during the initial start-up or ramp of the
specimen.
Check Furnace Alarm, Shutdown:
A fatal temperature event precedes this message. This indicates the
actual shutdown of the furnace control due to that event.
Control T/C Switched to Highest T/C:
Informational alarm which indicates that the temperature control is
completely stable and the control is now on the highest physical thermocouple in the furnace.
Furnace Split Exceeded:
Indicates the system has adjusted the power split between the zones as much as
possible and was not able to control temperature uniformity.
ISO Thermal Temperature Range Error, Shutdown:
The temperature measurement subsystem ambient
temperature sensor is outside the range of normal operation, which is less than 32 degrees Fahrenheit
or greater than 120 degrees Fahrenheit.
Rezero of Extensometer Requested:
The system has detected that the specimen has creeped close to the
range of the displacement transducer. The system is requesting a mechanical rezero of the extensometer.
Shutdown Due to Furnace Out of Band Time Exceeded:
The furnace has been out of band limits for
more than fifteen minutes and the ramp up will be terminated. Check for the following conditions:
1. Open furnace element or solid state relay
2. Specified ramp rate is too fast
3. Shorted thermocouple
4. Low AC line voltage
Specimen Broke During Loading:
The system was loading the specimen and after five minutes the break
switch never went to a not broken state. The system assumes that the specimen broke.
Specimen is Ready to Load Weights:
The specimen is ready to be loaded on manual frames.
Specimen Loaded:
The specimen was loaded by the system.
Содержание 2140-M Series
Страница 1: ...WinCCS Lever Arm Tester Series 2140 M 2320 M 2330 M 2330 CC 2410 M 2410 CC...
Страница 2: ...Revised March 2013...
Страница 14: ...14...
Страница 95: ...95 Once these steps are completed the thermocouples will be properly attached to the specimen...
Страница 109: ...109 Appendix C Default Values This appendix includes all the default values typically seen on a WinCCS tester...
Страница 110: ...110...
Страница 111: ...111...
Страница 112: ...112...
Страница 113: ...113...
Страница 115: ...154 East Brook Lane Butler PA 16002 USA 1 724 283 1212 www atspa com...