30
The WinCCS Modular requires the address to be set on the CPU Board via dip switches.
Each frame controller services two frames, and they have been designated left (main) and right
(secondary). The controller is commonly mounted on the main tester. Enter the proper designation in
the Side pull-down list under Frame Controller Information.
Except in the rare case that a frame controller is configured to support only one frame, the proper
designation under Type should be listed as Dual.
6.2.3 Frame Load Capacity
The minimum and maximum frame loads are specified by the frame manufacturer, and it is important
that the correct information be entered. The frame load specified during any test is checked against this
information, and tests which require loads outside the specified limits are not allowed to run. This can
protect the frame and load components from damage. To access frame load, click on System—Frame
Setup and enter the appropriate information under Frame Load Capacity.
NOTE: Do not change the minimum load capacity, or the test will lose accuracy.
6.2.4 Frame Naming
Each frame must be given a unique name. The default name, which is the frame number, can also be
used. To assign a specific name to a frame, go to Frame Setup. Click on the General Information box
and enter the frame title in the Name box. The name entered in this space will be viewed on the status
display screen and used in all reports and graphs.
At this point, the operator can now set up the various submenus listed under Frame Setup by clicking
through the tabs.
Содержание 2140-M Series
Страница 1: ...WinCCS Lever Arm Tester Series 2140 M 2320 M 2330 M 2330 CC 2410 M 2410 CC...
Страница 2: ...Revised March 2013...
Страница 14: ...14...
Страница 95: ...95 Once these steps are completed the thermocouples will be properly attached to the specimen...
Страница 109: ...109 Appendix C Default Values This appendix includes all the default values typically seen on a WinCCS tester...
Страница 110: ...110...
Страница 111: ...111...
Страница 112: ...112...
Страница 113: ...113...
Страница 115: ...154 East Brook Lane Butler PA 16002 USA 1 724 283 1212 www atspa com...