49
Frame Load Calibration/Verification:
This section allows the option of calibrating the frame with a load
cell rather than the traditional proving ring.
To ignore the tare measurement listed for the load cell on non-load control frames, check the box. To
review load cell information, select Load Cell(s) under the Frame tab on Calibration.
The Calibration and Verification recall periods allow the user to set the amount of time to pass before
the next extensometer calibration.
NOTE: Frames must be recalibrated annually. If a frame is not recalibrated, it will finish a currently-
running test on the frame, but no new tests can begin. Call ATS at +1-724-283-1212 to schedule
recalibration. Be sure to call in advance to ensure the procedure is scheduled in enough time.
6.3.7 Test Lockouts
This section prevents the operator from starting tests when various system components have not been
calibrated. All blocks in this section must be checked. Consult ATS at +1-724-283-1212 to schedule
recalibration.
6.3.8 Application Types
When creating a test specification, an application type is required. Under the Applications tab, add
or subtract specific applications from the list of entries in this field. Once a test is being created, the
operator can then pull from this list, ensuring uniformity of data across tests and test frames.
6.3.9 Site Information
The site information refers to the name of the lab that conducts the testing. This will appear on all
system reports. Enter the appropriate information under the Site Information tab.
Содержание 2140-M Series
Страница 1: ...WinCCS Lever Arm Tester Series 2140 M 2320 M 2330 M 2330 CC 2410 M 2410 CC...
Страница 2: ...Revised March 2013...
Страница 14: ...14...
Страница 95: ...95 Once these steps are completed the thermocouples will be properly attached to the specimen...
Страница 109: ...109 Appendix C Default Values This appendix includes all the default values typically seen on a WinCCS tester...
Страница 110: ...110...
Страница 111: ...111...
Страница 112: ...112...
Страница 113: ...113...
Страница 115: ...154 East Brook Lane Butler PA 16002 USA 1 724 283 1212 www atspa com...