Appendix A: System Status Display
A.1 Status Display Headings
Frame:
Each frame has an identifier, either a name assigned during the setup process, or a number from
one to the total system complement. The default lists frames in numerical order.
State:
The frame state allows the operator to monitor the process of the test.
Specimen Name:
The specimen name is the unique identifier assigned to the particular test subject
installed in the frame. This identifier will be used for all archival purposes for subsequent test data.
Setpoint:
The setpoint is the temperature specified in the test specification to which the specimen will
be tested. In case of room temperature tests and idle frames, N/A is displayed to indicate the furnace is
not in use.
Average:
This column displays the average of the temperatures of all the thermocouples in use on this
frame. The user thermocouple is not included in this average.
Top (Middle, Bottom):
The display here assumes that three thermocouples are in use, and the
temperature of the uppermost thermocouples is displayed here. The middle and lower thermocouples
are likewise shown in the adjacent columns.
Power:
At any time during the test, the furnace control algorithm will be varying the power into the
furnace in order to keep the temperature at the specified value. This display, which is a % of the total
power available, provides an indication to the operator of the power level being delivered to the furnace.
Time:
Depending on the frame state, time may have different meaning. For example, if a test is running
normally, time will be the total run time of the test, starting from initial specimen loading. A frame in
soak state will show the time remaining until specimen loading.
% Creep:
When a creep test is in process, this display will show the amount of creep of the specimen in
real time.
Alarms:
This display shows an abbreviation name for the alarm state of the frame. In addition, this
column indicates to the operator that some action at the individual frame is required. For example, load
weights, zero extensometer, etc. may appear in this column.
Содержание 2140-M Series
Страница 1: ...WinCCS Lever Arm Tester Series 2140 M 2320 M 2330 M 2330 CC 2410 M 2410 CC...
Страница 2: ...Revised March 2013...
Страница 14: ...14...
Страница 95: ...95 Once these steps are completed the thermocouples will be properly attached to the specimen...
Страница 109: ...109 Appendix C Default Values This appendix includes all the default values typically seen on a WinCCS tester...
Страница 110: ...110...
Страница 111: ...111...
Страница 112: ...112...
Страница 113: ...113...
Страница 115: ...154 East Brook Lane Butler PA 16002 USA 1 724 283 1212 www atspa com...