89
10.
Calibration Procedure
10.1.
Calibration Initialization
Press the 0 and 1 keys simultaneously and then turn the input power switch on. The program
will automatically enter to the calibration mode and the display screen will show:
Calibration Mode
: Forward
: Backward
Use the
(Forward) and
(Backward) keys to move through the calibration menu. The
following instructions follow the sequential Forward scrolling through the menu.
10.2.
Security Settings
10.2.1.
Password Setting
Press the
key, the program will advance to the Password setting mode. The display will
show:
Password = 0
Range : 0 - 9 9 9 9
or
Password = X X X X
Range : 0 - 9 9 9 9
The Password can be any four (4) digit number. If the Password is set to “0”, the keyboard
lock out will be selected by the LOCK key on the front panel without a Password. The
Password default is preset to “0” at the factory.
10.2.2.
Memory Lock selection
Press the
key, the program will advance to the Memory Lock selection mode. The display
will show:
MR-Lock = O N
ENTER
to Select
or
MR-Lock = O F F
ENTER
to Select
If the MR-Lock is selected ON, the Memory selection Menu will be disabled when in Lock
Mode. If the MR-Lock is selected OFF, the Memory selection Menu will be enabled when in
Lock Mode. Different memories can be recalled but the test parameters or steps can not be
changed. The MR-Lock default is preset to ON at the factory.