25
3.2.
Instrument Controls
3.2.1.
Front Panel Controls
1.
ARC LED:
This indicator will illuminate when the Arc Detector has detected an
arcing condition. This indicator will function even when the Arc Fail has been
disabled.
2.
LOCK LED:
When the Lock LED is on, the "password" software lockout has been
enabled. This means that the users will be unable to access the "program" mode of the
instrument to change any settings. When Memory-Lock is OFF the user can change
memory locations.
3.
BUS REMOTE LED:
This indicator will light when the instrument is in the Remote
Control mode. When the BUS Remote is on the instrument is able to send and
receive signals across the GPIB IEEE-488 or USB/RS-232 bus.
4.
LCD DISPLAY:
The 2x20 character display indicates test function, memory and
step location, test parameter and failure type as well as test measurements during a
test.
5.
POWER SWITCH:
Rocker-style switch with international ON (
) and OFF (0)
markings.
6.
RESET BUTTON:
This is a momentary contact switch. If an out-of-range leakage
current condition, an arc breakdown, HI-limit, LO-Limit or ground bond failure occur
you will need to reset the system before you can proceed to the next test. Press and
release the red RESET button.
ASSOCIATED
RESEARCH, INC.
R
500VA DIELECTRIC ANALYZER
7700
1
10 11
13
12
14 15
16
17
18 19 20
21
2 3
4
5
6
7
8 9