If Step 8 is set to ON the test process will be connected to the first step of the next
Memory.
4.2.3.
Insulation Resistance Test Setting
Press the INSULATION key and the display will show:
IR Set XXX.X s
MXX-X XXXX V XXXXX M
Note: X = the numeric
IR Set
: Insulation Resistance test setting screen
XXX.X s
: Delay Time setting
MXX
: Memory Program number
-X
: Test Step number
XXXX V
: DC Output Voltage
XXXX M
: Low-Limit of Insulation Resistance
Use the
or
arrow keys to progress through the test parameters menu. The
key will
advance forward and
key will advance backward. The sequential forward menu items
are Voltage, Charge-LO, HI-Limit, LO-Limit, Delay Time, Scanner Set, Connect.
Insulation Resistance Output Voltage setting
Advance the menu to the Voltage parameter. The display will show:
Voltage = XXXX V
Range : 0 – 6000
Use the Numeric keys to enter the voltage setting, then press the ENTER key. The
program will store the voltage setting and advance to the Charge-LO parameter
automatically. The unit is volt and 1 volt per step. Press the EXIT key to exit from the
setting mode to the operation mode if all parameters have been set.