Smart GFI allows the user to automatically configure the instruments return
configuration. When the Return lead is earth grounded, the GFI circuit is disabled and
the instrument operates in a grounded return mode of operation. Grounded Return allows
the user to perform tests on devices that have their chassis earth grounded by the test
fixture or test environment. The standard configuration of the instruments is a floating
return connection that is not directly connected to Earth ground. The standard
configuration allows monitoring of very low level leakage current without internal or
external stray earth ground leakage currents being measured and thereby causing errors in
the reading. Grounding the return will create some amount of leakage current that is
proportional to the High Voltage output and can cause small amounts of error depending
on the test voltage and physical environment.
5.6.
Reviewing Test Results
The test results from the first 8 test that are executed can be reviewed by pressing the
Enter/Review key. Successive key presses will continue advancing to the next result.
The results of the last step in the process will be followed by the first step when scrolling
through the results.
If two or more memories are connected together (more than an 8 step process) then only
the first 8 steps executed in the process will be stored into the results memory. The
results from the steps executed after the first 8 steps in the process will not be retrievable.
Results can be reviewed at any time before the next test is executed. All 8 buffers are
cleared at the start of the next test cycle.
5.7.
Displayed Messages
5.7.1.
Ground Bond Displayed Messages
(Model 7704 only)
If the test in process is Aborted with the RESET button or remote control, the display will
show:
GND Abort
XXX.X s
MXX-X XX.XX A XXX m
If the test in process is Aborted with the RESET button or remote control before the
meter readings are taken, the display will show:
GND Abort XXX.X s
MXX-X - - . - - A - - - m
or
GND Abort XXX.X s
MXX-X XX.XX A - - - m
At the beginning of the Ground Bond test, before the meter readings are taken, the
display will show: