73
Display Reading
Results Reading
DATA CODES
ASCII
DELIMITER
CR + LF (+ EOI)
8.7.
GPIB Address
Each device on the GPIB (IEEE-488) interface must have a unique address. You can set
the address of the 7704 or 7700 to any value between 0 and 30. The address is set to 8
when the instrument is shipped from the factory. The address can only be set from the
front panel. The address is stored in non-volatile memory and does not change when the
power has been off or after a remote interface reset.
8.8.
USB/RS-232 Interface
This interface provides all of the function control of the GPIB interface with the
exception of the SRQ functions. All commands can be found in the command list portion
of the Operation section 8.9 of this manual.
The USB/RS-232 cabling should be configured as follows for a 9 pin serial port interface:
The COM port should have the following configuration. 9600 baud, 8 data bits, 1 stop
bit, no parity. This interface does not support XON/XOFF protocol or any hardware
handshaking. The controller should be configured to ignore the handshaking lines DTR
(pin 4), DSR (pin 6) CTS (pin 8) and RTS (pin 7). If the port can not be configured
through software to ignore the lines then the handshake lines should be jumpered together
in two different sets. Pins 4 and 6 jumpered together and pins 7 and 8 jumpered together
at the controller end of the cable.
When sending command over the USB/RS-232 bus the instrument will echo a response
string identical to the string that was sent if the transfer was recognized and accepted by
the instrument. This allows for software handshaking, to monitor and control data flow.
If there is an error with the command string that is sent, the instrument will respond with
15h or the “NAK” ASCII control code. When requesting data from the instrument, it
automatically send the data to the controller input buffer. The controller input buffer will
Instrument RS-232 Port
PC / Bus Controller
2
3
5
2
3
5
TD
RD
TD
RD
SIG
GND
SIG
GND