If the DUT current exceeds the HI-Limit of Insulation Resistance test and the resistance
is not within the metering range, the display will show:
IR HI-Limit
XXX.X s
MXX-X XXXX V
9999 M
If the DUT current falls below the LO-Limit of Insulation Resistance test and the
resistance is within the metering range, the display will show:
IR LO-Limit
XXX.X s
MXX-X XXXX V XXXX M
If the DUT current falls below the LO-Limit of Insulation Resistance test and the
resistance is below the metering range, the display will show:
IR LO-Limit
XXX.X s
MXX-X XXXX V
1M
If the GFI threshold is exceeded during of the Insulation Resistance test, the display will
show:
IR GND-Fault XXX.X s
MXX-X XXXX V XXXX M
When the DUT passed the Insulation Resistance test, when the test process is complete
the display will show:
IR Pass
XXX.X s
MXX-X XXXX V XXXX M