The output power supplies of this instrument are referenced directly to
earth ground. Any conductor that completes a path between the high
voltage and earth ground will form a completed circuit.
Return Connection
When the instrument Return is grounded, any internal and external stray leakage will be
monitored due to currents that flow from High Voltage to earth ground (such as from HV to
the chassis of the instrument). This current is inherent and will cause errors when trying to
monitor very low leakage currents in the microamp range.
Safe Lifting and Carrying Instructions
Proper methods of lifting and carrying can help to protect against injury. Follow the
recommendations below to ensure that instruments are handled in a safe manner.
Determine if the instrument can be lifted by one individual or requires additional
support.
Make sure that your balance is centered and your feet are properly spaced, shoulder
width apart behind the instrument.
Bend at the knees and make sure your back is straight.
Grip the instrument with your fingers and palms and do not lift unless your back is
straight.
Lift up with your legs, not your back.
Keep the instrument close to your body while carrying.
Lower the instrument by bending your knees. Keep you back straight.
2.2.4.
Environmental Conditions
This equipment is intended for indoor use only. The equipment has been evaluated according
to Installation Category II and Pollution Degree 2 as specified in IEC 664.
This instrument may be operated in environments with the following limits:
Temperature………….41° - 104° F (5° - 40° C)
Relative humidity …….0 - 80%
Altitude ………………6,560 feet (2,000 meters)
Keep the ventilation slits uncovered during operation. Failure to do
so could cause the instrument to overheat and may damage internal
components.