52
4.3.
System Parameter Settings
Use the SETUP key to progress through the menu of System Parameters. Successive key
presses will advance the menu forward. The sequential forward menu items are: PLC
Remote, Address, Contrast, Volume, Fail Stop.
The setting of system parameters affect the operating conditions of the instrument and are
separate from the functional settings. The system settings are also global and are not
specific to any memory location.
PLC Remote selection
Press the Setup key to advance to the PLC Remote parameter. The display will show:
PLC Remote = ON
ENTER
to Select
or
PLC Remote = OFF
ENTER
to Select
Use the ENTER key to select the mode PLC Remote Control.
After selecting PLC Remote mode, press the SETUP key to advance to the address
setting if the GPIB interface card is installed on this instrument. If this instrument does
not have the GPIB interface card, the program will advance to the contrast setting, or
press the EXIT key to exit from the PLC Remote selection to the operation mode. The
instrument will store the selection of PLC Remote automatically.
If the Remote Control is set to ON, the test function will be controlled by the Remote
Control via the remote connectors located on the rear panel. The TEST button on the
front panel is disabled but the RESET button is still enabled.
The remote Memory Program recall functions can be performed only when the PLC
Remote is set ON. In addition, when the PLC remote is set to ON the remote TEST
signal input is active while in the Bus Remote Mode.
If the Remote Control is set to OFF, the operation of the instrument will be controlled by
the local TEST and RESET buttons on the front panel.
GPIB Address setting
If the GPIB interface card is not installed on this instrument, this setup program will not
appear on the setup menu.
Press the Setup key to advance the menu to the address parameter. The display will
show:
Address =
XX