Fail to Stop/Continue selection
Press the Setup key to advance the menu to the Fail Stop parameter. The display will
show:
Fail Stop = ON
ENTER
to Select
or
Fail Stop = OFF
ENTER
to Select
Use the ENTER key to select the mode of Fail Stop.
After the selection of Fail Stop mode, press the SETUP key to advance to the beginning
of the menu (PLC Remote), or press the EXIT key to exit from the Fail Stop selection to
the operation mode. The Program will store the selection of Fail Stop automatically.
This function is used when multiple steps have been connected. If the Fail Stop mode is
set ON, the test process will stop at the step that fails, or at the end of test process. When
there are unfinished steps to be completed, pressing the TEST button will continue the
test process at the next step. Pressing the RESET button and then the TEST button will
start the test process from the first step.
If the Fail Stop is set to OFF, the instrument will continue to complete the entire test
process even if there were failed steps during the test. The display will show the results
of each step at the end of test process, as follows:
Test P
F
P F
P
F
P F
P
F
P F P
F
P
F
Step 1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
First Memory Program
Second Memory Program
The pass fail status is indicated by the letter F (Fail) or P (Pass). The Pass/Fail status of
the first memory Program will show on the left end and the second one will be on the
right. The pass/fail status of the first 16 steps can be displayed. A maximum of 16 steps
will be saved. If more than 16 steps are executed in sequence. Results from those steps
greater than 16 cannot be retrieved from memory.