19
3.
Specifications and Controls
3.1.
Model 7704 and 7700
Functional Specifications
INPUT
Voltage
100/115/200/230 VAC
10
, Single Phase, User selection
Frequency
47 - 63 Hz
Fuse
15 Amp 250V fast blow internal
DIELECTRIC WITHSTAND TEST MODE
Output Rating
5 KV @ 100 mA AC,
6 KV @ 10 mA DC
Output Adjustment
Range:
0 - 5 KV AC, 0 - 6 KV DC
Resolution: 1 volt/step
Accuracy:
(2
of setting
5 volts)
Ramp-HI
12mA peak maximum, ON/OFF selectable
Charge-LO
Range:
0.0 - 350.0
A DC or Auto set
HI-Limit AC Range:
0.00 - 99.00 mA
Resolution: 0.01 mA/step
DC Range:
0 - 9999
A
Resolution: 1
A/step
Accuracy: AC or DC
(2
of setting
2 counts)
LO-Limit AC Range:
0.000 - 9.999 mA
Resolution: 0.001 mA/step
DC Range:
0.0 - 999.9
A
Resolution: 0.1
A/step
Accuracy: AC or DC
(2
of setting
2 counts)
Arc Detection
Range:
1 - 9
Failure Detector
Audible and Visual
Voltage Display
Range:
0.00 - 6.00 KV Full Scale
Resolution: 10 volt/step
Accuracy:
(2
of reading
2 counts)
Содержание 7700
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 30: ...22 Ground Fault Interrupt GFI Trip Current 450 A max AC or DC HV Shut Down Speed 1 ms ...