95
G-Voltage = V
Enter Standard V-out
Use the Numeric keys to enter the standard value of voltage, unit V, and then press the ENTER
key to store the standard value of AC Ground Bond voltage and display will show:
Ground V, 7V OK
T E S T
t o C a l i b r a t e
10.3.12.
Calibration of Ground Bond AC Current (Model 7704 only)
Press the
key, the program will advance to the AC Current calibration of the Ground bond
test. The display will show:
Ground A, 30A
T E S T
t o C a l i b r a t e
Connect a AC standard Ammeter which can measure up to 30AAC to the High current and
Return output leads, then press the TEST button, the program will automatically generate an
output on the output leads of about 30AAC and the display will show:
G- Current = A
Enter Standard I-out
Use the Numeric keys to enter the standard value of current, unit A, and then press the ENTER
key to store the standard value of AC Ground Bond current and display will show:
Ground A, 30A OK
T E S T
t o C a l i b r a t e
10.3.13.
Calibration of Real Current for units with option 51 only
This menu item will be shown only when the Real Current option is installed.
Press the
key, the program will advance to the Real Current calibration of the AC
withstand mode. The display will show:
AC 5mAR, 500K
T E S T
t o C a l i b r a t e
Connect a resistor about 500K
/10W in series with a AC standard Ammeter which can
measure up to 10mA to the output leads. The Ammeter should be connected to the return lead,
then press the TEST button, the program will automatically generate an output of about
5000V/10mA and the display will show:
AC Current = mA
Enter Standard I-out