4.1.2.
Memory Configuration
Each memory location contains 8 separate steps which can be connected sequentially to
the next consecutive step. Each step contains 4 separate locations, 1 for each function.
Only one function can be selected for each step, but the other three function parameters
remain stored in the memory. The following memory map illustrates the separate
locations for each function.
Memory map
Memory 1
Step 1
Step 2
. . . .
Step 8
ACW
ACW
ACW
DCW
DCW
DCW
IR
IR
IR
G-Bond
G-Bond
G-Bond
Memory 2
Step 1
Step 2
. . . .
Step 8
ACW
ACW
ACW
DCW
DCW
DCW
IR
IR
IR
G-Bond
G-Bond
G-Bond
Memory 50
Note: On model 7700 disregard the location for G-Bond parameters. The Continuity
parameter is stored within the ACW and DCW functions as a single setting for both
modes.
4.1.3.
Memory Program Selection
Press the MEMORY key, the display will show:
Memory = X X
Range : 1 - 50
Use the Numeric Key to enter the Memory location number, and then press the ENTER
key. The program will recall the test parameters stored in this location and return to the
operation mode automatically.