Chapter IV Local Operation
62
ARRAY
4.5.1 CC Mode
CC mode has two ranges: the high range (CCH) and the low range (CCL).
Example 1:
Take 3751A as an example, in CVL mode, set the load as CCH mode, the current
level is 80A. Turn on the load. The procedures are shown as below:
Step 1:
Press
key to enter into
mode and parameter setting menu;
Step 2:
Press dynamic
key to select
mode;
Step 3:
Enter the current level 80.000 by the number keypad, or via knob with
, and
keys;
Step 4:
Press
key to confirm;
Step 5:
Switch to other optional via
and
keys;
Step 6:
Press
key to open the load.
The corresponding SCPI commands of the above procedures should be as follows:
MODE
CCH
; Set the mode
CURR
80
; Set the current level
INP
ON
; Turn on the load
Example 2:
In CCH mode, set the current level to 16A, the current rising rate to 0.1A/us, the
current falling rate to 0.5A/us, the VON voltage to 10V and the current limit level 25A. Turn on
the load. The procedures are shown as below:
Step 1:
Switch to parameter setting menu via
and
keys;
Step 2:
Move the cursor to the ten position via the
key, then enter number 1 via the knob or
the number keypad; move the cursor to the digit position via the
key, then enter number 6
via the knob or the number keypad; then press
key to confirm;
Step 3:
Switch to Volt Start option via
and
keys;
Step 4:
Move the cursor to the ten position via the
key, then enter number 1 via the knob or
Содержание 3750A
Страница 1: ...Contents USER S MANUAL 375X SERIES ELECTRONIC LOAD ...
Страница 63: ...Chapter III Installation 59 ARRAY ...
Страница 114: ...Chapter V Remote Programming Operation 110 ARRAY Frame Diagram Fig 1 ...
Страница 115: ...Chapter V Remote Programming Operation 111 ARRAY ...
Страница 116: ...Chapter V Remote Programming Operation 112 ARRAY Fig 2 ...
Страница 120: ......